AS 200 Air Jet Sieve Machine


The Air Jet Sieving Machine AS 200 jet is particularly suitable for low density and low particle size materials which tend to agglomerate. It is used with sieves of 10 microns mesh size and up. The procedure is very gentle on the material as no mechanical sieving aids are required. The average sieving time is only 2-3 minutes.

The AS 200 jet is specifically designed for test sieves with a diameter of 203 mm/8“ (or 200 mm with adapter). The air jet generated by an industrial vacuum cleaner can be adjusted by using the manual vacuum regulation. Optionally, an automatic vacuum regulation is available. The Open Mesh Function, a procedure which greatly reduces the number of near-mesh particles, provides optimum separation efficiency, excellent reproducibility and a longer service life of the sieves. Sieving time and nozzle speed are conveniently selected with a single button; the settings are shown in the graphic display. The Quick Start Mode is used to start the sieving process under standard conditions without entering parameters.

AS 200 JET SIEVE MACHINE INTRODUCTION

Accuracy & Efficiency

  • Air jet technology for dispersion and deagglomeration
  • Measuring range 10 μm to 4 mm
  • Quick, efficient procedure
  • Open Mesh Function reduces near-mesh particles
  • Digital parameter setting (time, vacuum, speed)
  • Quick Start option
  • Variable nozzle speed
  • Automatic vacuum regulation and cyclone (options)
  • Memory for 9 Standard Operating Procedures (SOPs)
  • Suitable for RETSCH standard sieves

ACCESSORIES

RETSCH sieve shakers can be equipped with a variety of accessories to meet a wide range of application requirements.

Cyclone with holder and collecting receptable

To extend the service life of the filters in the vacuum cleaner and for recovery of the sample fraction passing the sieve, we recommend the use of the optional cyclone. The separation degree and limiting particle size respectively depend on the sample characteristics.

Test sieves 10 μm and up

RETSCH test sieves consist of a solid stainless steel sieve frame of high stability for reliable sieving results. Paying close attention to mesh-specific requirements, the fabric of each test sieve is precisely joined into the frame and tautened.

Ultrasonic baths and dryers

Suitable for thorough cleaning of test sieves and for quick, gentle drying of samples and sieves.

Sample Dividers

Meaningful results can only be obtained if the sample represents the original material. Sample dividers produce representative part samples, thus ensuring reproducibility of the analysis.

EASYSIEVE / CFR EVALUATION SOFTWARE

EasySieve, the software for particle size analyses, exceeds manual evaluation in many aspects. The software is able to automatically control the necessary measurement and weighing procedures – from the registration of the weight of the sieve up to the evaluation of the data. It is simple and convenient to use and is also available in an FDA 21 CFR Part 11-conform version.

TYPICAL SAMPLE MATERIALS

Vibratory sieve shakers are frequently used for grain size analysis of cement clinker, chemicals, coffee, construction materials, fertilizers, fillers, flours, grains, metals powders, minerals, nuts, plastics, sand, seeds, soils, washing powder, etc.

Coffee
Soil
Cereals
Construction Materials

FUNCTIONAL PRINCIPAL

The AS 200 jet is equipped with a rotating slotted nozzle. The sieve and lid are placed above the nozzle. A vacuum generates a strong jet of air which disperses the particles on the sieve through the slotted nozzle. The material which is smaller than the mesh size of the sieve is transported by the back flow of the air to the cyclone or to the vacuum cleaner. The air jet helps to deagglomerate the particles and constantly purges the sieve mesh.

TECHNICAL DATA

Applications separation, fractioning, particle size determination
Field of application chemistry / plastics, construction materials, environment / recycling, food, geology / metallurgy, glass / ceramics, medicine / pharmaceuticals
Feed material powders
Measuring range* 10 µm – ~ 4 mm
Sieving motion dispersion by air jet
Max. batch / feed capacity 0.3 – 100 g
Max. number of fractions 1 (by using a cyclone: 2)
Speed digital, 5 – 55 min-1 (nozzle)
Time display digital, 00:10 – 99:59 min
Interval operation no
Vacuum 2,000 – 9,999 Pa / 20 – 99 mbar / 0.3 – 1.45 psi
Storable SOPs 9 / Quick Start mode
Suitable for dry sieving yes
Suitable for wet sieving no
Serial interface yes
Including test certificate / can be calibrated yes
Suitable sieve diameters 203 mm (200 mm with adapter)
Max. height of sieve stack 1 sieve (25 mm (1″) / 50 mm ( 2″))
Accessories cyclone / automatic vacuum regulation
Electrical supply data 100-240 V, 50/60 Hz
Power connection 1-phase
W x H x D 460 x 288 x 305 (without cover)
Net weight ~ 16 kg
Standards CE

*depending on feed material and instrument configuration/settings